@article{oai:rakuno.repo.nii.ac.jp:00001699, author = {森, 勝伸 and 廣永, 貴大 and 佐鳥, 達也 and 板橋, 英之 and NAKATANI, Nobutake and 中谷, 暢丈 and 小崎, 大輔 and 田中, 一彦}, issue = {5}, journal = {鉄と鋼}, month = {May}, note = {Article, Simultaneous chromatographic separation of nitrate ion (NO_3^−), nitrite ion (NO_2^−) and ammonium ion (NH_4^+) was obtained using single injection, single column and a detector (conductivity). The principle is based on selective ion-exclusion effect for NH_4^+ and anion-exchange effects for NO_3^− and NO_2^− on an OH^−-formed basic anion-exchange resin column (Tosoh TSKgel Super IC-AZ) with basic eluent (2 mol/m^3 KOH). The present system could determine NH_4^+, NO_2^− and NO3− in 500-fold diluted artificial seawater sample in 40 min at 0.6 mL/min of flow rate, by standard addition method. Also, this method was applicable to nitrogen analysis in artificial seawater with iron matrix, by pre-column packed with an iminodiacetate chelating resin in order to remove iron or calcium ions.}, pages = {273--278}, title = {アンモニウム,亜硝酸,硝酸イオンのイオン排除/陰イオン交換クロマトグラフィーによる同時定量の開発および高塩濃度試料への応用}, volume = {97}, year = {2011} }