@article{oai:rakuno.repo.nii.ac.jp:00006429, author = {NISHI, Yasunobu and MURAKAMI, Yoshiki and OTSUKA, Marina and TSUKANO, Kenji and AJITO, Tadaharu and 鈴木, 一由}, issue = {9}, journal = {Journal of Veterinary Medical Science}, month = {Sep}, note = {Article, The aim of this study was to evaluate the accuracy and precision of portable blood glucose meters, such as i-STAT 1 and Precision Xceed, for use in calves. Whole blood and plasma samples were obtained from eleven calves that received 2.5 or 5.0% dextrose-containing polyelectrolyte isotonic solutions. Measurements using the i-STAT 1 (r2=0.99, P<0.0001) and Precision Xceed (r2=0.96, P<0.0001) were well correlated with those by the hexokinase method, which is the gold standard. Although the accuracy of i-STAT 1 was equivalent to that of the hexokinase method, there was an autocorrelation in the residuals between the results from the Precision Xceed and the hexokinase method. Thus, the i-STAT 1 can be used to measure the blood glucose concentration in cattle.}, pages = {1301--1304}, title = {Accuracy of the point-of-care glucose meter for use in calves}, volume = {81}, year = {2019} }